Home / How to read a peptide HPLC chromatogram and peak table

How to read a peptide HPLC chromatogram and peak table

Find the integration basis, peak assignments and sample identifiers behind a reported HPLC area percentage.

Read the axes before reading the percentage

Locate the time axis, response axis, units and detector designation. Then find the peak table and the trace that generated it. Waters defines peak area using the curve and its baseline. The area column and the peak height column describe different measurements. If the report highlights a main peak, find the row assigned to it and check that the sample and acquisition identifiers match the plotted trace.

Waters · HPLC terminology and interpretation

Red-capped sample vials sit beneath the head of an HPLC autosampler.
Capped HPLC sample vialsVial positions provide a physical point of reference when matching a sample identifier to an analytical record.
Image creditsMartin Schaier · Image source · CC BY 4.0 · Converted to WebP.
A liquid chromatography instrument and a separate mass spectrometer share a laboratory work area.
Chromatography and mass spectrometry instrumentationThe reference instrument photograph gives context for the identity report requested from the actual testing laboratory.
Image creditsSarka Na kopci · Image source · CC BY-SA 4.0 · Converted to WebP.

Identify what contributes to the total

The meaning of a normalized area percentage depends on the integrated signals included in its denominator. Ask the issuing laboratory which peaks were counted, whether any were excluded under the named method and how exclusions are recorded. An unexplained small peak should remain an observation until its identity is supported. Keep the full plotted time range and legend, especially when an excerpt is being used in a supplier presentation.

Bachem · Quality control of amino acids and peptides

Resolve an apparent mismatch between the graph and table

If a graph seems to show three peaks while the table has two rows, mark the particular feature on a copy and return the original file identifier to the laboratory. Ask whether the feature was integrated, deliberately excluded or outside the reported channel. Request the laboratory’s written explanation or revised attachment. Keep the earlier file with the response so the purchasing reviewer can understand why the reported percentage was accepted or questioned.

Bachem · Quality control of amino acids and peptides

Worked example

Two chromatogram exports for one sample may show different time windows. Ask the laboratory to identify which trace and integration table belong to the issued result and mark the displayed interval on each. A purchasing reviewer can then choose a figure without omitting documented context.

Prepare your inquiry

An interpretable purity figure belongs to a named sample, trace and integration basis.

Select the items you want to ask about.

Certificates of analysis

For lyophilized peptide powders, match the material name, lot and report number before comparing the COA with the chromatogram and mass report.

Read a COA