Identify the limit the decision will use
An analytical number and a pass or fail statement serve different purchasing decisions. If a buyer wants the laboratory to assess conformity, send the exact specification, its revision, the relevant property and reporting unit. Ask who is entitled to interpret that requirement. ILAC G8 discusses decision rules and statements of conformity, including the importance of understanding the rule behind a conclusion. Record whether the laboratory offers this statement in the quoted service.


Image credits
Giovanna Canu, Francesca Cirisano, Sofia Gambaro, CNR-ICMATE · Image source · CC BY 4.0 · Converted to WebP.Agree what happens near the limit
In an illustrative example, a specification sets a minimum of 95.0 units and a result is displayed as 95.0 after rounding. That display alone leaves open which underlying value is evaluated and how measurement uncertainty is treated. Ask the laboratory to explain the decision rule in terms your procurement reviewer can use. Put the accepted wording and responsible reviewer in the service brief before anyone has to interpret a borderline report.
Keep the statement within the examined scope
A report might support a conformity statement for one specified property while reporting other properties for information. Ask the laboratory to name the samples, results and specification clauses its statement covers. If several lots were submitted, request a separate traceable conclusion for each relevant lot or a clear description of the grouping. Keep your own final purchasing decision with the report so future reviewers can distinguish the laboratory’s statement from the buyer’s release decision.
Prepare your inquiry
Request a conformity statement only after the evaluator, specification and decision rule are clear.